Other features:
*Internal temperature compensation sensor RIC, temperature range -10~50℃. The temperature measurement accuracy at 18~28°C is ±0.5°C, and the other temperature measurement accuracy is ±1°C. The cold junction compensation time is 10 times/time. * Output maximum applied voltage: about 32Vpk: output maximum applied current: about 5mA
*Output protection: 50mA/60V composite film insurance
*Temperature coefficient: 0.1×basic accuracy/°C (temperature range <18°C or >28°C)
Measurement function [applicable within one year after approval, 23°C±5°C, 35~70%RH, accuracy (% set value + word]
Other features:
*Internal temperature compensation sensor RIC, temperature range -10~50℃. The temperature accuracy of 18~28°C is ±0.5°C, and the other color temperature accuracy is ±1°C. The cold junction compensation time is 105/time.
* Display refresh rate: 2 times/second.
*The maximum applied voltage at the input terminal: 30Vpk.
* Input Common Mode Rejection: 50Hz/60Hz>120dB: Input Series Mode Rejection: 50Hz/60Hz>60dB
*Temperature coefficient: 0.1×basic accuracy/℃
(temperature range <18°C or >28°C)